- 显 示:
- 240*64 点阵液晶蓝屏
- 检测原理:
- 卡尔费休库仑滴定(电量滴定)
- 测量范围:
- 0.1μg~100mg
The trace moisture meter is designed using the Karl Fischer Coulomb Method (electricity method) *Blue screen LCD display Full Chinese menu Human-machine interface Friendly instrument The built-in calculation formula can be displayed directly The moisture content percentage detection circuit of the sample uses digital processing technology to overcome the temperature drift of traditional measuring instruments and greatly improve the accuracy and sensitivity of the test. The whole machine adopts a unique relay-less process design to completely avoid the hidden dangers caused by oxidation of memory relay contacts and extend the service life of the instrument. Users with a printer interface can choose a micro-printer to save data as needed. It has been widely used in power, petrochemical, pharmaceutical and other departments and is one of the most cost-effective products launched by our company.
_
_
_ 1
Technical Parameters
Display * Dot Matrix LCD Blue Screen
Measurement Detection Constant Current Polarization Detection
Detection Principle Karl Fischer Coulomb titration (coulometric titration)
Self-checking electrode faults are automatically detected and displayed
Calculation formulas are built-in for commonly used calculation formulas, volume injection, weighing injection, direct display of water content
Measuring range
Sensitivity
Accuracy
(excluding sampling error)
The clock will not lose the year, month, day, week, hour, minute and second when the clock is powered off
Applicable ambient temperature ℃
Applicable ambient humidity
Power
Dimensions**
_