- 温度:
- 20℃±2℃
- 湿度:
- 95%RH以上
Technical requirements for a concrete curing room
The temperature standard of the concrete curing room is ℃, and the relative humidity standard is above. The uniformity of indoor temperature and humidity should meet the above requirements
There should be a rack for placing specimens in the room. , the specimens are placed on the shelf at a distance of at least~
No matter which humidification device is used, it must be ensured that the sprayed water is in an atomized state, and water cannot be poured directly on the specimens
2. Concrete Inspection method of curing room
Direct observation inspection should meet the requirements of placing the test pieces on the shelf and keeping the distance between each other at least ~. It must be ensured that the sprayed water is in an atomized state, and water cannot be poured directly on the test pieces.
Use a ventilated wet and dry bulb thermometer to measure the temperature and humidity at each point in the standard curing room. Measure once every hour for a total of times
The measuring points are arranged as follows. Within the actual use range, at least five measuring points should be arranged. , that is, measuring points are set according to the upper and lower diagonals of the room, a point is set at a corner indoors, and a measuring point is set at the intersection of the two diagonals.
General equipment of three concrete curing rooms
Test block placement rack
Constant temperature and humidity controller (one set)
Drainage ditch
Automatic atomizing nozzle
Indoor Insulation treatment is done around the wall
Four Intelligent Concrete Curing Rooms
With the development of technology, standard curing room equipment has been systematic, intelligent, and automated.
Configuration of five concrete curing rooms
Nanjing Jiedi Electronic Instrument Equipment Manufacturing Co., Ltd.
Temperature host (industrial variable frequency refrigeration unit, with cooling, heating and dehumidification functions)
Humidity sensor (imported Temperature and humidity sensing module)
Industrial-grade ultrasonic humidifier
System control box
Screen LCD display temperature controller (with communication interface, can be connected to a computer printer)
Air and return air system
Multipoint
Technical parameters of six concrete curing rooms
Nanjing Jiedi Electronic Instrument Equipment Manufacturing Co., Ltd.
Model
Power distribution
Power
Cooling
Output
Power
Cooling (heat) amount
Control
Area
㎡
Humidity
Temperature
( ℃)
Water source
Power supply
Refrigeration
Amount
Heating capacity
Heat
Pump
Electricity
Heat
℃
℃
Tap water
Three Phase
Five lines
< br/>
Remarks
According to customer requirements, we can configure an all-in-one machine or a split machine. The performance of the two models is exactly the same. The split machine is suitable for customers with limited space around underground test rooms or curing rooms.
Nanjing Jiedi Electronic Instrument Equipment Manufacturing Co., Ltd.